Features
-
Users can save measurement result data and custom methods directly in spreadsheet mode
-
32GB of storage space is pre-installed
-
Integrated fiber to prevent fibre breakage due to external force collision, which results in chaotic measurement results
-
There is no need to warm up, start testing whenever you want
-
Built-in Win10 system, built-in 7-inch high-definition display, full touch operation
-
Stainless steel and quartz fiber is used for the platform
-
Using a liquid column to identify abnormal function
-
The built-in Bluetooth and WIFI allow for remote operation and wireless printing of results
-
Two USB ports that can connect to a variety of devices such as mice, keyboards, desktop computers, and so on
Specifications
|
Wavelength Range |
190 to 850 nm |
|
Wavelength Accuracy |
± 1nm |
|
Wavelength Resolution |
2 nm (FWHM at Hg 546 nm) |
|
Repeatability |
<± 0.5 nm |
|
Light Path |
1 mm, 0.05 mm (optical path automatic conversion) |
|
Absorbance Accuracy |
1% (0.76 absorbance at 350 nm) |
|
Absorbance Range |
0.002 to 300Abs (equivalent to 10mm path length) |
|
sample volume Requirements |
0.3 to 2 μl |
|
Cuvette Specifications |
1 mm, 2 mm, 5 mm, 10 mm |
|
Min sample volume of the cuvette |
20 μl |
|
The minimum height of the cuvette sample |
5 mm |
|
Temperature Control |
4 to 42 ℃ thermostat adjustable temperature error <± 0.5 ℃ |
|
Nucleic Acid measurement range |
0.2 to 37500 mg / μl (sa DNA) |
|
Protein measurement range |
0.01 to 1120 mg / ml (BSA) |
|
Sample measurement time |
less than 3s |
|
Sample base material |
stainless steel and quartz fiber |
|
Light Source |
long life pulsed xenon flash lamp |
|
Detector |
2048(CMOS) element linear silicide CCD array |
Application
Nucleic acid, protein, bacteria, cell bacteria, microarray samples, and conventional full-wavelength scanning, among other things, can be detected.